Assessment of antibiotic aerosol generation using commercial jet nebulizers

J Aerosol Med. 1994 Fall;7(3):217-28. doi: 10.1089/jam.1994.7.217.

Abstract

The performance of 14 commercial jet nebulisers has been assessed; Unineb, Suremist (Unimed (UK) Ltd), Micro-Cirrus (Intersurgical Ltd), Pulmo-Neb (DeVilbiss Health Care UK Ltd), Side-Stream (Medic-Aid Ltd), Micro-Neb III (Lifecare Ltd), RespirGard (Marquest Medical Products Inc), Aeromist, Venticaire (S and W Vickers Ltd), Up-Draft II, Ava-Neb, Up-Draft (Hudson Respiratory Care Inc), Bennett/Twin, Raindrop (Puritan-Bennett Corporation). The units were operated with a high flow compressor (Maxi III, Medix Ltd) at 101/min. Performance was assessed by measuring the fraction of the initial mass of drug released as an aerosol and nebulisation time for initial drug volume of 2-6mls, and the mass median diameter and mass fraction of the aerosol in particles < 5.17 microns diameter. The Side-Stream nebuliser gave the best performance, although incorporation of a filter to trap exhaled antibiotic may prove difficult. The Micro-Cirrus generated a particularly fine aerosol. The Raindrop nebuliser performed well, while the Up-Draft II nebulised efficiently but was associated with extended nebulisation times which may limit its utility.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Administration, Inhalation
  • Anti-Bacterial Agents / administration & dosage*
  • Anti-Bacterial Agents / pharmacology
  • Evaluation Studies as Topic
  • Kinetics
  • Nebulizers and Vaporizers*

Substances

  • Anti-Bacterial Agents