Recognition of impending ketoacidosis delayed by ketone reagent strip failure

JAMA. 1978 Nov 24;240(22):2462-4.

Abstract

During a one-year period, six patients were admitted to the hospital in diabetic ketoacidosis following 12 to 24 hours of illness at home. Urine testing with strips impregnated with sodium nitroprusside had indicated that no ketone bodies were present. Nine of 11 bottles of reagent strips in use for less than three months by these and other patients contained totally unreactive strips. Nitroprusside tablets exposed more than three months continued to give accurate readings.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Diabetic Ketoacidosis / diagnosis*
  • Diagnostic Errors
  • Health Education
  • Home Nursing
  • Humans
  • Ketone Bodies / urine*
  • Nitroprusside
  • Product Labeling
  • Reagent Kits, Diagnostic* / standards

Substances

  • Ketone Bodies
  • Reagent Kits, Diagnostic