False alarms in very low birthweight infants: comparison between three intensive care monitoring systems

Acta Paediatr. 2000 May;89(5):571-6. doi: 10.1080/080352500750027880.

Abstract

Monitor alarms are a major burden on both patients and staff in intensive care units. We compared alarm rates from three different monitor systems (Hewlett Packard (HP), Kontron Instruments (KI), Marquette-Hellige (MH)) in a tertiary neonatal intensive care unit. Monitors were used in random order on three consecutive days over 8 h each in 16 preterm infants (median gestational age at birth 29 wk (range 24-34), age at study 18 d (8-53), weight at study 1,160g (595-1,430)). Alarms were classified as true or false using flow sheets based on continuous observation of both the patient and related parameters. There was one alarm every 9 min of monitoring. The median number of true alarms did not differ significantly between systems, being 28 per 8 h (range 9-87) for HP, 26 (3-81) for KI, and 30 (5-135) for MH. The median number of false alarms differed widely, with the HP system generating 32 (7-77) such alarms per 8 h, compared to 8 (0-19) for KI and 15 (2-32) for MH (p < 0.01 HP vs KI and MH, p < 0.05 KI vs MH). These differences between systems were mainly due to differences in pulse oximeter and transcutaneous PO2 monitor alarm rates.

Conclusions: In conclusion, this study shows marked differences between both parameters and manufacturers in the frequency with which false alarms occur. It may provide a basis from which reductions in alarm rates can be sought.

Publication types

  • Clinical Trial
  • Comparative Study
  • Controlled Clinical Trial

MeSH terms

  • Child Health Services / organization & administration*
  • False Positive Reactions
  • Germany
  • Humans
  • Hypoxia / diagnosis*
  • Infant
  • Infant, Newborn
  • Infant, Very Low Birth Weight*
  • Intensive Care Units, Neonatal / organization & administration*
  • Monitoring, Physiologic / instrumentation*
  • Oximetry / methods
  • Predictive Value of Tests