Determination of the enantiomers of ketamine and norketamine in human plasma by enantioselective liquid chromatography-mass spectrometry

J Chromatogr B Analyt Technol Biomed Life Sci. 2003 Aug 25;794(1):99-108. doi: 10.1016/s1570-0232(03)00420-3.

Abstract

A sensitive enantioselective liquid chromatographic assay with mass spectrometric detection has been developed and validated for the simultaneous determination of plasma concentrations of (R)- and (S)-ketamine, and (R)- and (S)-norketamine. The compounds were extracted from human plasma using solid-phase extraction and then directly injected into the LC-MS system for detection and quantification. Enantioselective separations were achieved on a liquid chromatographic chiral stationary phase based upon immobilized alpha(1)-acid glycoprotein (the Chiral AGP column). The separations were achieved using a mobile phase composed of 2-propanol-ammonium acetate buffer (10 mM, pH 7.6) (6:94, v/v), a flow-rate of 0.5 ml/min and a temperature of 25 degrees C. Under these conditions, the analysis time was 20 min. Detection of the ketamine, norketamine and bromoketamine (internal standard) enantiomers was achieved using selected ion monitoring at m/z 238.1, 224.1 and 284.0, respectively. Extracted calibration curves were linear from 1 to 125 ng/ml per enantiomer for each analyte with correlation coefficients better than 0.9993 and intra- and inter-day RSDs of less than 8.0%. The method was applied to samples from a clinical study of ketamine in pain management.

MeSH terms

  • Calibration
  • Chromatography, Liquid / methods*
  • Humans
  • Ketamine / analogs & derivatives*
  • Ketamine / blood*
  • Mass Spectrometry / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Stereoisomerism

Substances

  • Ketamine
  • norketamine