A comparison of algorithms for body-worn sensor-based spatiotemporal gait parameters to the GAITRite electronic walkway

J Appl Biomech. 2012 Jul;28(3):349-55. doi: 10.1123/jab.28.3.349. Epub 2011 Nov 14.

Abstract

This study compares the performance of algorithms for body-worn sensors used with a spatiotemporal gait analysis platform to the GAITRite electronic walkway. The mean error in detection time (true error) for heel strike and toe-off was 33.9 ± 10.4 ms and 3.8 ± 28.7 ms, respectively. The ICC for temporal parameters step, stride, swing and stance time was found to be greater than 0.84, indicating good agreement. Similarly, for spatial gait parameters--stride length and velocity--the ICC was found to be greater than 0.88. Results show good to excellent concurrent validity in spatiotemporal gait parameters, at three different walking speeds (best agreement observed at normal walking speed). The reported algorithms for body-worn sensors are comparable to the GAITRite electronic walkway for measurement of spatiotemporal gait parameters in healthy subjects.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Exercise Test / instrumentation*
  • Female
  • Gait / physiology*
  • Humans
  • Male
  • Monitoring, Ambulatory / instrumentation*
  • Monitoring, Ambulatory / methods*
  • Pattern Recognition, Automated / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Transducers, Pressure*
  • Young Adult