Detection of silicone oil leakages in freeze dryers

PDA J Pharm Sci Technol. 2011 Sep-Oct;65(5):481-5. doi: 10.5731/pdajpst.2011.00748.

Abstract

As a small leakage from the silicone oil tubes in pharmaceutical freeze dryers may go undetected for several cycles, it provides a major threat to both product safety and the economic performance of such installations. A properly mounted mass spectrometer offers a non-invasive method of detecting the smallest amounts of silicone oil down to 1 ppm. Additionally the system will improve leak testing procedures and can also be used as a process analytical technology tool to improve process control.

Lay abstract: Drugs, and in particular parenterals, need to have a very high level of purity. A production process such as freeze drying carries the inherent risk of contamination by process fluids. This should not happen during routine production, but in the event of a malfunction small amounts of these fluids may contaminate the drug product. The purpose of this paper is to introduce a method which is able to detect such malfunctions before they can develop into a risk for drug purity and ultimately patient safety.