Note: Development of 9 A current source for precise resistance measurement method

Rev Sci Instrum. 2012 Oct;83(10):106105. doi: 10.1063/1.4759022.

Abstract

In this Note, design of voltage controlled current source intended to be used in precise resistance measurement system in the range from 0.1 mΩ to 10 Ω is presented. The design specifications of current source include gross-tuning of current in the range from 0.1 mA to 9 A, low noise, low temperature coefficient, and short term stability better than 50 ppm. The realized current source has achieved better short term stability than comparable commercial devices.