Polysomnographic validation of a wireless dry headband technology for sleep monitoring in healthy young adults

Physiol Behav. 2013 Jun 13:118:185-8. doi: 10.1016/j.physbeh.2013.05.036. Epub 2013 May 26.

Abstract

The present study aimed to explore the validity and reliability of a wireless dry headband technology for sleep monitoring (WS), through a comparison with concurrent polysomnographic (PSG) recording in healthy young adults. Eleven volunteers (7 females; mean age±SD: 24.75±3.62years) took part in the study, wearing the WS for two overnight PSG recordings in the sleep laboratory. The WS was compared to PSG in the identification of wake, light, deep and REM sleep. The WS sensitivity and specificity were 97.6% and 56.1%, respectively. The WS agreement with PSG, measured by Cohen's kappa, was 0.56 for light sleep, 0.70 for deep sleep, and 0.67 for REM sleep. The present results showed that the agreement ranged from moderate to high between PSG and the WS, while wakefulness detection was observed to be a limitation of the WS.

Keywords: Ambulatory; Automated; Polysomnography; Reliability; Sleep; Wireless dry headband technology.

Publication types

  • Validation Study

MeSH terms

  • Data Interpretation, Statistical
  • Female
  • Humans
  • Male
  • Polysomnography / instrumentation*
  • Reproducibility of Results
  • Sleep / physiology*
  • Sleep Stages / physiology
  • Sleep, REM / physiology
  • Wireless Technology*
  • Young Adult