Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience

Eur Radiol. 2014 Aug;24(8):1889-95. doi: 10.1007/s00330-014-3209-7. Epub 2014 May 17.

Abstract

Objectives: To evaluate image quality, maximal heart rate allowing for diagnostic imaging, and radiation dose of turbo high-pitch dual-source coronary computed tomographic angiography (CCTA).

Methods: First, a cardiac motion phantom simulating heart rates (HRs) from 60-90 bpm in 5-bpm steps was examined on a third-generation dual-source 192-slice CT (prospective ECG-triggering, pitch 3.2; rotation time, 250 ms). Subjective image quality regarding the presence of motion artefacts was interpreted by two readers on a four-point scale (1, excellent; 4, non-diagnostic). Objective image quality was assessed by calculating distortion vectors. Thereafter, 20 consecutive patients (median, 50 years) undergoing clinically indicated CCTA were included.

Results: In the phantom study, image quality was rated diagnostic up to the HR75 bpm, with object distortion being 1 mm or less. Distortion increased above 1 mm at HR of 80-90 bpm. Patients had a mean HR of 66 bpm (47-78 bpm). Coronary segments were of diagnostic image quality for all patients with HR up to 73 bpm. Average effective radiation dose in patients was 0.6 ± 0.3 mSv.

Conclusions: Our combined phantom and patient study indicates that CCTA with turbo high-pitch third-generation dual-source 192-slice CT can be performed at HR up to 75 bpm while maintaining diagnostic image quality, being associated with an average radiation dose of 0.6 mSv.

Key points: • CCTA is feasible with the turbo high-pitch mode. • Turbo high-pitch CCTA provides diagnostic image quality up to 73 bpm. • The radiation dose of high-pitch CCTA is 0.6 mSv on average.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Artifacts*
  • Coronary Angiography / methods*
  • Coronary Artery Disease / diagnostic imaging*
  • Coronary Artery Disease / physiopathology
  • Female
  • Heart Rate
  • Humans
  • Imaging, Three-Dimensional*
  • Male
  • Middle Aged
  • Multidetector Computed Tomography / methods*
  • Phantoms, Imaging*
  • Prospective Studies
  • Radiation Dosage
  • Reproducibility of Results