Finger Prick Dried Blood Spots for HIV Viral Load Measurement in Field Conditions in Zimbabwe

PLoS One. 2015 May 22;10(5):e0126878. doi: 10.1371/journal.pone.0126878. eCollection 2015.

Abstract

Background: In the context of a community-randomized trial of antiretrovirals for HIV prevention and treatment among sex workers in Zimbabwe (the SAPPH-IRe trial), we will measure the proportion of women with HIV viral load (VL) above 1000 copies/mL ("VL>1000") as our primary endpoint. We sought to characterize VL assay performance by comparing results from finger prick dried blood spots (DBS) collected in the field with plasma samples, to determine whether finger prick DBS is an acceptable sample for VL quantification in the setting.

Methods: We collected whole blood from a finger prick onto filter paper and plasma samples using venipuncture from women in two communities. VL quantification was run on samples in parallel using NucliSENS EasyQ HIV-1 v2.0. Our trial outcome is the proportion of women with VL>1000, consistent with WHO guidelines relating to regimen switching. We therefore focused on this cut-off level for assessing sensitivity and specificity. Results were log transformed and the mean difference and standard deviation calculated, and correlation between VL quantification across sample types was evaluated.

Results: A total of 149 HIV-positive women provided DBS and plasma samples; 56 (63%) reported being on antiretroviral therapy. VL ranged from undetectable-6.08 log10 using DBS and undetectable-6.40 log10 using plasma. The mean difference in VL (plasma-DBS) was 0.077 log10 (95%CI = 0.025-0.18 log10; standard deviation = 0.63 log10,). 78 (52%) DBS and 87 (58%) plasma samples had a VL>1000. Based on plasma 'gold-standard', DBS sensitivity for detection of VL>1000 was 87.4%, and specificity was 96.8%.

Conclusion: There was generally good agreement between DBS and plasma VL for detection of VL>1000. Overall, finger prick DBS appeared to be an acceptable sample for classifying VL as above or below 1000 copies/mL using the NucliSENS assay.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Specimen Collection*
  • Desiccation
  • Female
  • HIV Infections / virology*
  • HIV-1*
  • Humans
  • Sensitivity and Specificity
  • Serologic Tests*
  • Specimen Handling
  • Viral Load*
  • Zimbabwe

Grants and funding

This research was funded by United Nations Population Fund (UNFPA) Zimbabwe with pooled funding from the United Kingdom Department for International Development (DFID), the Government of Sweden, and Irish Aid. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.