Competitive Protein-binding assay-based Enzyme-immunoassay Method, Compared to High-pressure Liquid Chromatography, Has a Very Lower Diagnostic Value to Detect Vitamin D Deficiency in 9-12 Years Children

Int J Prev Med. 2015 Jul 27:6:67. doi: 10.4103/2008-7802.161069. eCollection 2015.

Abstract

Background: The most reliable indicator of Vitamin D status is circulating concentration of 25-hydroxycalciferol (25(OH) D) routinely determined by enzyme-immunoassays (EIA) methods. This study was performed to compare commonly used competitive protein-binding assays (CPBA)-based EIA with the gold standard, high-pressure liquid chromatography (HPLC).

Methods: Concentrations of 25(OH) D in sera from 257 randomly selected school children aged 9-11 years were determined by two methods of CPBA and HPLC.

Results: Mean 25(OH) D concentration was 22 ± 18.8 and 21.9 ± 15.6 nmol/L by CPBA and HPLC, respectively. However, mean 25(OH) D concentrations of the two methods became different after excluding undetectable samples (25.1 ± 18.9 vs. 29 ± 14.5 nmol/L, respectively; P = 0.04). Based on predefined Vitamin D deficiency as 25(OH) D < 12.5 nmol/L, CPBA sensitivity and specificity were 44.2% and 60.6%, respectively, compared to HPLC. In receiver operating characteristic curve analysis, the best cut-offs for CPBA was 5.8 nmol/L, which gave 82% sensitivity, but specificity was 17%.

Conclusions: Though CPBA may be used as a screening tool, more reliable methods are needed for diagnostic purposes.

Keywords: Competitive protein-binding assays; Vitamin D deficiency; Vitamin D measurement; high-pressure liquid chromatography.