Nasal Expiratory Positive Airway Pressure Devices (Provent) for OSA: A Systematic Review and Meta-Analysis

Sleep Disord. 2015:2015:734798. doi: 10.1155/2015/734798. Epub 2015 Dec 21.

Abstract

Objective. To quantify the effectiveness of nasal expiratory positive airway pressure (nasal EPAP) devices or Provent as treatment for obstructive sleep apnea (OSA). Methods. PubMed and six other databases were searched through November 15, 2015, without language limitations. Results. Eighteen studies (920 patients) were included. Pre- and post-nasal EPAP means ± standard deviations (M ± SD) for apnea-hypopnea index (AHI) in 345 patients decreased from 27.32 ± 22.24 to 12.78 ± 16.89 events/hr (relative reduction = 53.2%). Random effects modeling mean difference (MD) was -14.78 events/hr [95% CI -19.12, -10.45], p value < 0.00001. Oxygen desaturation index (ODI) in 247 patients decreased from 21.2 ± 19.3 to 12.4 ± 14.1 events/hr (relative reduction = 41.5%, p value < 0.00001). Lowest oxygen saturation (LSAT) M ± SD improved in 146 patients from 83.2 ± 6.8% to 86.2 ± 11.1%, MD 3 oxygen saturation points [95% CI 0.57, 5.63]. Epworth Sleepiness Scale (ESS) M ± SD improved (359 patients) from 9.9 ± 5.3 to 7.4 ± 5.0, MD -2.5 [95% CI -3.2, -1.8], p value < 0.0001. Conclusion. Nasal EPAP (Provent) reduced AHI by 53.2%, ODI by 41.5% and improved LSAT by 3 oxygen saturation points. Generally, there were no clear characteristics (demographic factors, medical history, and/or physical exam finding) that predicted favorable response to these devices. However, limited evidence suggests that high nasal resistance could be associated with treatment failure. Additional studies are needed to identify demographic and polysomnographic characteristics that would predict therapeutic success with nasal EPAP (Provent).

Publication types

  • Review