Pilot study of the multicentre DISCHARGE Trial: image quality and protocol adherence results of computed tomography and invasive coronary angiography

Eur Radiol. 2020 Apr;30(4):1997-2009. doi: 10.1007/s00330-019-06522-z. Epub 2019 Dec 16.

Abstract

Objective: To implement detailed EU cardiac computed tomography angiography (CCTA) quality criteria in the multicentre DISCHARGE trial (FP72007-2013, EC-GA 603266), we reviewed image quality and adherence to CCTA protocol and to the recommendations of invasive coronary angiography (ICA) in a pilot study.

Materials and methods: From every clinical centre, imaging datasets of three patients per arm were assessed for adherence to the inclusion/exclusion criteria of the pilot study, predefined standards for the CCTA protocol and ICA recommendations, image quality and non-diagnostic (NDX) rate. These parameters were compared via multinomial regression and ANOVA. If a site did not reach the minimum quality level, additional datasets had to be sent before entering into the final accepted database (FADB).

Results: We analysed 226 cases (150 CCTA/76 ICA). The inclusion/exclusion criteria were not met by 6 of the 226 (2.7%) datasets. The predefined standard was not met by 13 of 76 ICA datasets (17.1%). This percentage decreased between the initial CCTA database and the FADB (multinomial regression, 53 of 70 vs 17 of 75 [76%] vs [23%]). The signal-to-noise ratio and contrast-to-noise ratio of the FADB did not improve significantly (ANOVA, p = 0.20; p = 0.09). The CTA NDX rate was reduced, but not significantly (initial CCTA database 15 of 70 [21.4%]) and FADB 9 of 75 [12%]; p = 0.13).

Conclusion: We were able to increase conformity to the inclusion/exclusion criteria and CCTA protocol, improve image quality and decrease the CCTA NDX rate by implementing EU CCTA quality criteria and ICA recommendations.

Key points: • Failure to meet protocol adherence in cardiac CTA was high in the pilot study (77.6%). • Image quality varies between sites and can be improved by feedback given by the core lab. • Conformance with new EU cardiac CT quality criteria might render cardiac CTA findings more consistent and comparable.

Keywords: Angiography, coronary; CT angiography; Coronary artery disease; Medical imaging; Trial protocols.

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Computed Tomography Angiography / methods*
  • Coronary Angiography / methods*
  • Coronary Artery Disease / diagnosis*
  • Coronary Vessels / diagnostic imaging*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pilot Projects
  • Reproducibility of Results

Grants and funding