Development and validation of an ultra-high performance liquid chromatography with tandem mass spectrometry method for the simultaneous quantification of direct oral anticoagulants in human plasma

J Chromatogr B Analyt Technol Biomed Life Sci. 2021 Oct 1:1182:122952. doi: 10.1016/j.jchromb.2021.122952. Epub 2021 Sep 21.

Abstract

Direct oral anticoagulants are widely used to treat and prevent thromboembolic disorders. With rising clinical application, monitoring concentrations of direct oral anticoagulants are necessary in certain clinical conditions. A rapid and sensitive ultra-performance liquid chromatography-tandem mass spectrometry method was developed for the simultaneous determination of dabigatran etexilate, dabigatran, rivaroxaban, edoxaban, and apixaban, in human plasma. Protein precipitation with methanol was performed for sample preparation. The direct oral anticoagulants and internal standards were separated under gradient conditions using a C18 column, at an analytical run time of 8 min. The mobile phase was composed of 0.1% (v/v) formic acid in water (solvent A) and 0.1% (v/v) formic acid in acetonitrile (solvent B) at a flow rate of 0.3 mL/min. Mass detection was performed in multiple reaction monitoring using positive ionization mode. The method was validated over a range of 1.0-500 ng/mL for dabigatran etexilate, 0.1-500 ng/mL for dabigatran, and 0.5-500 ng/mL for edoxaban, rivaroxaban, and apixaban. The method detection limits of five analytes were in the range of 0.05-0.5 ng/mL. The lower limits of quantification of five analytes ranged from 0.1 to 1 ng/mL. The linearity (r2 values) was higher than 0.997. The accuracy of the low, medium, and high quality control samples were between 85.9 and 114%, and intra- and inter-day precision were below 9.47%. This validated method was successfully used to determine the plasma concentrations of rivaroxaban in 32 patients, and of dabigatran etexilate and dabigatran in 1 patient.

Keywords: DOAC; Human plasma; Simultaneous determination; Therapeutic drug monitoring; UHPLC-MS/MS.

MeSH terms

  • Anticoagulants / blood*
  • Chromatography, High Pressure Liquid / methods*
  • Drug Monitoring
  • Humans
  • Limit of Detection
  • Linear Models
  • Reproducibility of Results
  • Tandem Mass Spectrometry / methods*

Substances

  • Anticoagulants