High-Sensitivity and High-Throughput Quantification of Everolimus in Human Whole Blood Using Ultrahigh-Performance Liquid Chromatography Coupled With Tandem Mass Spectrometry

Ther Drug Monit. 2022 Oct 1;44(5):633-640. doi: 10.1097/FTD.0000000000000985.

Abstract

Background: Rigorous dose adjustment by therapeutic drug monitoring (TDM) is recommended when everolimus (EVR) is administered for immunosuppression. In this study, the authors developed a highly sensitive ultrahigh-performance liquid chromatography coupled with the tandem mass spectrometry (UHPLC-MS/MS) method for measuring EVR concentrations in whole blood using a high-throughput solid-phase extraction method for sample pretreatment. Furthermore, the blood EVR concentrations in routine TDM samples from patients who underwent renal transplantation measured using the established UHPLC-MS/MS method were compared with those measured using the latex agglutination turbidimetric immunoassay (LTIA).

Methods: Blood samples were pretreated by solid-phase extraction using a 96-well HLB µElution plate. The clinical application of the newly developed method was evaluated using 87 blood samples from 19 patients who underwent kidney transplant.

Results: The calibration curve showed good linearity over a wide range of 0.1-50 ng/mL, with relative error ≤15% obtained from the back calculation of calibrators, and ≤20% for the lower limit of quantification. Within-batch and batch-to-batch accuracies and precisions fulfilled the acceptance criteria of the US Food and Drug Administration guidelines for bioanalytical method validation. The extraction recovery rates were good (≥65.2%), and almost no matrix effects were found in any of the quality control samples. Blood EVR concentrations measured by UHPLC-MS/MS were positively correlated with those measured by LTIA. A Bland-Altman plot indicated that the UHPLC-MS/MS method yielded better measurements than the LTIA method, regardless of the concentration.

Conclusions: Therefore, the authors succeeded in developing a novel high-sensitivity and high-throughput method for measuring blood EVR concentration by UHPLC-MS/MS using a µElution plate for sample pretreatment.