Simultaneous determination of diquat, paraquat, glufosinate, and glyphosate in plasma by liquid chromatography/tandem mass spectrometry: from method development to clinical application

Anal Bioanal Chem. 2024 May;416(12):3073-3083. doi: 10.1007/s00216-024-05257-1. Epub 2024 Mar 22.

Abstract

Diquat (DQ), paraquat (PQ), glufosinate (GLU), and glyphosate (GLYP) are commonly used herbicides that have been confirmed to be toxic to humans. Rapid and accurate measurements of these toxicants in clinical practice are beneficial for the correct diagnosis and timely treatment of herbicide-poisoned patients. The present study aimed to establish an efficient, convenient, and reliable method to achieve the simultaneous quantification of DQ, PQ, GLU, and GLYP in human plasma using liquid chromatography-tandem mass spectrometry (LC-MS/MS) without using derivatization or ion-pairing reagents. DQ, PQ, GLU, and GLYP were extracted by the rapid protein precipitation and liquid-liquid extraction method and then separated and detected by LC-MS/MS. Subsequently, linearity, limit of detection (LOD), limit of quantification (LOQ), precision, accuracy, extraction recovery, matrix effect, dilution integrity, and stability were evaluated to validate the method based on the FDA criteria. Finally, the validated method was applied to real plasma samples collected from 166 Chinese patients with herbicide poisoning. The results showed satisfactory linearity with low LOD (1 ng/mL for DQ and PQ, 5 ng/mL for GLU, and 10 ng/mL for GLYP, respectively) and low LOQ (5 ng/mL for DQ and PQ, 25 ng/mL for GLU and GLYP, respectively). In addition, the precision, accuracy, extraction recovery, and stability of the method were acceptable. The matrix effect was not observed in the analyzed samples. Moreover, the developed method was successfully applied to determine the target compounds in real plasma samples. These data provided reliable evidence for the application of this LC-MS/MS method for clinical poisoning detection.

Keywords: Diquat; Glufosinate; Glyphosate; Human plasma; LC-MS/MS; Paraquat.

Publication types

  • Validation Study

MeSH terms

  • Aminobutyrates* / blood
  • Chromatography, Liquid / methods
  • Diquat* / blood
  • Diquat* / poisoning
  • Glycine* / analogs & derivatives
  • Glycine* / blood
  • Glyphosate*
  • Herbicides* / blood
  • Herbicides* / poisoning
  • Humans
  • Limit of Detection*
  • Paraquat* / blood
  • Paraquat* / poisoning
  • Reproducibility of Results
  • Tandem Mass Spectrometry* / methods

Substances

  • Glycine
  • phosphinothricin
  • Glyphosate
  • Aminobutyrates
  • Diquat
  • Paraquat
  • Herbicides