Determination of the glucocorticoid fluticasone propionate in plasma by automated solid-phase extraction and liquid chromatography-tandem mass spectrometry

J Chromatogr B Biomed Sci Appl. 1998 Nov 6;718(2):243-50. doi: 10.1016/s0378-4347(98)00374-0.

Abstract

A sensitive, robust and high throughput mass spectrometry based method is described for the determination of the glucocorticoid fluticasone propionate in plasma. The method employs solid-phase extraction in 96 well microtitre plate format which has been automated by means of a custom built Zymark robotic system. The extracts are analysed by liquid chromatography-tandem mass spectrometry using thermally and pneumatically assisted electrospray ionisation and selected reaction monitoring. The method is both accurate and precise with both intra- and inter-assay precision (C.V.) of less than <6%. The method provides a lower limit of quantification of 20 pg/ml from 0.5 ml of human plasma, sufficient to monitor systemic concentrations of inhaled fluticasone propionate at therapeutic doses.

MeSH terms

  • Androstadienes / blood*
  • Anti-Asthmatic Agents / blood*
  • Anti-Inflammatory Agents / blood*
  • Chromatography, Liquid / methods*
  • Fluticasone
  • Humans
  • Mass Spectrometry / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Androstadienes
  • Anti-Asthmatic Agents
  • Anti-Inflammatory Agents
  • Fluticasone