Sample Multiplexing: Increased Throughput for Quantification of Total Testosterone in Serum by Liquid Chromatography-Tandem Mass Spectrometry

Clin Chem. 2020 Sep 1;66(9):1181-1189. doi: 10.1093/clinchem/hvaa117.

Abstract

Background: For high-volume assays, optimizing throughput reduces test cost and turn-around time. One approach for liquid chromatography-tandem mass spectrometry (LC-MS/MS) assays is sample multiplexing, wherein the analyte of interest is derivatized in different specimens with reagents of different molecular weight (differential mass tagging). Specimens can then be combined and simultaneously analyzed within a single injection to improve throughput. Here we developed and validated a quantitative, sample-multiplexed LC-MS/MS assay for serum total testosterone (TT) based on this approach.

Methods: For the sample-multiplexed assay, calibrators, controls, and patient specimens were first extracted separately. After mass tagging with either methoxyamine or hydroxylamine, they were combined and injected into the LC-MS/MS system. To evaluate assay performance, we determined limit of quantification (LOQ), linearity, recovery, and imprecision. A method-comparison study was also performed, comparing the new assay with the standard LC-MS/MS assay in 1574 patient specimens.

Results: The method was linear from 2.5 to 2000 ng/dL, with accuracies from 93% to 104% for both derivatives. An LOQ of 1.0 ng/dL was achieved. Intra-assay and total CVs across 4 quality control concentrations were less than 10%. The assay demonstrated good agreement (Deming regression, 1.03x + 6.07) with the standard LC-MS/MS assay for the patient specimens tested (TT, 3 to 4862 ng/dL).

Conclusion: Sample multiplexing by differential mass tagging of TT increases LC-MS/MS throughput 2-fold without compromising analytical accuracy and sensitivity.

MeSH terms

  • Chromatography, Liquid / methods*
  • Humans
  • Tandem Mass Spectrometry / methods*
  • Testosterone / blood*

Substances

  • Testosterone