Migration of the Duraloc cup at two years

J Bone Joint Surg Br. 1999 Jan;81(1):51-3. doi: 10.1302/0301-620x.81b1.9036.

Abstract

We carried out 71 primary total hip arthroplasties using porous-coated, hemispherical press-fit Duraloc '100 Series' cups in 68 consecutive patients; 61 were combined with the cementless Spotorno stem and ten with the cemented Lubinus SP II stem. Under-reaming of 2 mm achieved a press-fit. Of the 71 hips, 69 (97.1%) were followed up after a mean of 2.4 years. Migration analysis was performed by the Ein Bild Rontgen Analyse method, with an accuracy of 1 mm. The mean total migration after 24 months was 1.13 mm. Using the definition of loosening as a total migration of 1 mm, it follows that 30 out of 63 cups (48%) were loose at 24 months.