Evaluation of the first automated thyroglobulin assay

Clin Chem Lab Med. 1999 Feb;37(2):159-64. doi: 10.1515/CCLM.1999.028.

Abstract

The aim of this study was to investigate technical and analytical performance of the first automated thyroglobulin (Tg) assay (DPC-Immulite; Diagnostic Products Corporation, Los Angeles, USA). In imprecision studies using several human serum pools ranging from 21 to 58 replicates, a coefficient of variation of 9.0% was obtained at a mean Tg concentration of 0.84 ng/ml and of 6.1% at a Tg concentration of 62.1 ng/ml. In a method comparison with a non-automated assay (BRAHMS LUMItest Tg, BRAHMS, Berlin, Germany) using 383 sera of 303 patients with thyroid carcinoma, regression analysis according to Passing and Bablock yielded in the following equation: Immulite Tg = 1.6 x BRAHMS Tg-0.1 ng/ml (Pearson's r = 0.979). Sera obtained from 59 patients with thyroid carcinoma enabled comparative follow-up studies; in all cases qualitative agreement was found with regard to increase or decrease of serum Tg; in eight cases, however, Tg was detected with the Immulite assay but not with the BRAHMS assay. Further follow-up proved the presence of thyroid tissue in these patients. From these and further methodological data (dilution linearity, interference studies, carry-over study, high-dose hook properties, and short report time) it is concluded that the DPC-Immulite Tg assay meets the requirements of routine diagnostic use.

MeSH terms

  • Automation
  • Evaluation Studies as Topic
  • Humans
  • Immunoassay / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Thyroglobulin / blood*
  • Thyroid Diseases / blood
  • Thyroid Neoplasms / blood

Substances

  • Thyroglobulin