Abstract
To evaluate quantitative positioning errors of frameless stereotactic radiation therapy with a fusion of computed tomography (CT) and linear accelerator unit, Z-type CT markers were attached to patients, and CT images were obtained before and after daily treatment. In 40 verification tests, geometrical errors were never more than 1 mm.
Publication types
-
Research Support, Non-U.S. Gov't
MeSH terms
-
Brain Neoplasms / radiotherapy
-
Dose Fractionation, Radiation
-
Head / anatomy & histology
-
Humans
-
Phantoms, Imaging
-
Radiosurgery / instrumentation*
-
Radiotherapy Planning, Computer-Assisted
-
Radiotherapy, Conformal / methods*
-
Stereotaxic Techniques* / instrumentation
-
Tomography, X-Ray Computed* / instrumentation