Rapid and accurate quantitation of Hb Bart's and Hb H using weak cation exchange high performance liquid chromatography: correlation with the alpha-thalassemia genotype

Hemoglobin. 1999 Aug;23(3):203-11. doi: 10.3109/03630269909005700.

Abstract

The Bio-Rad Variant Hemoglobin testing system is an automated high performance liquid chromatography analyzer marketed with a beta-thalassemia short program to quantify Hbs F and A2, and assist in detecting Hbs A, S, D, C, and E. Although the two hemoglobins present in Hb H disease, Hb Bart's and Hb H, are separated by the system, they are not quantitated. In this study we modified the beta-thalassemia short program in order to facilitate quantitation of Hb Bart's and Hb H. Blood samples from 60 patients with Hb H disease, with various underlying genotypes, were studied. Analyses were performed on the day of blood collection or on hemolysates stored at -80 degrees C in cyanide or carbomonoxy forms. The mean sum of Hb Bart's and Hb H levels in all patients was found to be 12% (range 1.8-35%). Patients with nondeletional mutations (or association of alpha(0) deletion and nondeletional mutations) had notably higher Hb Bart's and Hb H levels when compared to patients with deletional genotypes.

MeSH terms

  • Adult
  • Chromatography, High Pressure Liquid / methods
  • Genotype
  • Globins / genetics
  • Hemoglobin H / analysis*
  • Hemoglobins, Abnormal / analysis*
  • Humans
  • Infant, Newborn
  • Isoelectric Focusing / methods
  • Mutation
  • Reproducibility of Results
  • alpha-Thalassemia / genetics
  • alpha-Thalassemia / metabolism

Substances

  • Hemoglobins, Abnormal
  • Globins
  • Hemoglobin H
  • hemoglobin Bart's