Paradigm shift in lead design

Pacing Clin Electrophysiol. 1999 Sep;22(9):1321-32. doi: 10.1111/j.1540-8159.1999.tb00625.x.

Abstract

During the past 30 years there has been a tremendous development in electrode technology from bulky (90 mm2) to pin-sized (1.0 mm2) electrodes. Simultaneously, impedance has increased from 110 Ohms to >1 kOhms, which has been termed a "paradigm shift" in lead design. If current is responsible for stimulation, why is its impedance a key factor in saving energy? Further, what mechanism is behind this development based on experimental findings and what conclusion can be drawn from it to optimize electrode size? If it is assumed that there is always a layer of nonexcitable tissue between the electrode surface and excitable myocardium and that the electric field (potential gradient) produced by the electrode at this boundary is reaching threshold level, then a formula can be derived for the voltage threshold that completely describes the electrophysiology and electrophysics of a hemispherical electrode. Assuming that the mean chronic threshold for porous steroid-eluting electrodes is 0.6 V with 0.5-ms pulse duration, thickness of nonexcitable tissue can be estimated to be 1.5 mm. Taking into account this measure and the relationship between chronaxie and electrode area, voltage threshold, impedance, and energy as a function of surface area can be calculated. The lowest voltage for 0.5-ms pulse duration is reached with r(o) = 0.5 d, yielding a surface area of 4 mm2 and a voltage threshold of 0.62 V, an impedance of 1 kOhms, and an energy level of 197 nJ. It can be deduced from our findings that a further reduction of surface areas below 1.6 mm2 will not diminish energy threshold substantially, if pulse duration remains at 0.5 ms. Lowest energy is reached with t = chronaxie, yielding an energy level <100 nJ with surface areas < or =1.5 mm2. It is striking to see how well the theoretically derived results correspond to the experimental findings. It is also surprising that the hemispheric model so accurately approximates experimental results with differently shaped electrodes that it can be concluded that electrode shape seems to play a minor role in electrode efficiency. Further energy reduction can only be achieved by reducing the pulse duration to chronaxie. A real paradigm shift will occur only if the fundamentals of electrostimulation in combination with electrophysics are accepted by the pacing community.

MeSH terms

  • Animals
  • Cardiac Pacing, Artificial / methods
  • Dogs
  • Electric Impedance
  • Electrodes, Implanted
  • Electrophysiology
  • Equipment Design
  • Pacemaker, Artificial*