Brain fluorine-18 fluorodeoxyglucose imaging with dual-head coincidence gamma camera: comparison with dedicated ring-detector positron emission tomography

AJNR Am J Neuroradiol. 2000 Jan;21(1):99-104.

Abstract

Background and purpose: Dual-head coincidence gamma camera (DHC) imaging has been proposed as an alternative to dedicated ring-detector positron emission tomography (dr-PET) for clinical fluorodeoxyglucose (FDG) studies. The purpose of this investigation was to assess the quality of DHC images in FDG studies of the human brain.

Methods: Seven healthy volunteers and 12 patients with various cerebral disorders underwent consecutive brain dr-PET and DHC with FDG. All sets of images were compared semiquantitatively using regions of interest.

Results: Cortical count ratios to the cerebellum on DHC and dr-PET images did not differ significantly among the volunteers, except in the superior frontal cortex and thalamus. In all studies including those of cerebral disorders, the mean cortical-to-cerebellar ratios of DHC and dr-PET images correlated closely.

Conclusion: FDG imaging with DHC delineated the metabolic distribution of glucose in the brain as well as dr-PET did, except in the superior frontal cortex and thalamus. Therefore, DHC may be a dedicated cost-effective means of detecting metabolic abnormalities in the brain.

Publication types

  • Clinical Trial
  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Brain Diseases / diagnostic imaging*
  • Female
  • Fluorodeoxyglucose F18*
  • Humans
  • Male
  • Middle Aged
  • Radiopharmaceuticals*
  • Tomography, Emission-Computed*

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18