The Kerboull acetabular reinforcement device in major acetabular reconstructions

Clin Orthop Relat Res. 2000 Sep:(378):155-68. doi: 10.1097/00003086-200009000-00025.

Abstract

Sixty consecutive revision total hip arthroplasties were performed with bulk allograft bone supported by the Kerboull reinforcement acetabular device in 53 patients from 1980 to 1987. The average age of the patients at the time of hip revision was 57.7 years. Acetabular bone loss according to the American Academy of Orthopaedic Surgeons grading system was Type III for 48 hips in 41 patients and Type IV for 12 hips in 12 patients. Three failures, defined as radiologic loosening of the socket, revised or not, were reported in this series at a mean 8-year followup. Eight patients died of unrelated causes at a mean of 5 years. No patient was lost to followup. The mean followup of the series was 10 years +/- 3 years. The mean preoperative Merle d'Aubigné hip functional score was 11.7 +/- 2.4 versus 17.4 +/- 0.6 at the latest followup. Consolidation of the graft was considered completed in all 60 hips and occurred by 12 months. Remodeling of the graft proceeded for 3 to 4 years. The survival rate at 13 years was 92.1% +/- 5% using loosening of the acetabular component as the end point. This study indicated that acetabular allograft reconstructions reinforced by the Kerboull acetabular device were able to provide satisfactory long-term clinical and radiologic results.

Publication types

  • Case Reports

MeSH terms

  • Acetabulum / surgery*
  • Adult
  • Aged
  • Aged, 80 and over
  • Arthroplasty, Replacement, Hip*
  • Disease Progression
  • Female
  • Humans
  • Male
  • Middle Aged
  • Osteoarthritis, Hip / surgery
  • Plastic Surgery Procedures
  • Prostheses and Implants*
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies