Abstract

We carried out weight-bearing video radiological studies on 40 patients with a total knee arthroplasty (TKA), to determine the presence and magnitude of femoral condylar lift-off. Half (20) had posterior-cruciate-retaining (PCR) and half (20) posterior-cruciate-substituting (PS) prostheses. The selected patients had successful arthroplasties with no pain or instability. Each carried out successive weight-bearing knee bends to maximum flexion, and the radiological video tapes were analysed using an interactive model-fitting technique. Femoral lift-off was seen at some increment of knee flexion in 75% of patients (PCR TKA 70%; PS TKA 80%). The mean values for lift-off were 1.2 mm with a PCR TKA and 1.4 mm with a PS TKA. Lift-off occurred mostly laterally with the PCR TKA, and both medially and laterally with the PS TKA. Separation between the femoral condyles and the articular surface of the tibia was recorded at 0 degrees, 30 degrees, 60 degrees and 90 degrees of flexion. Femoral condylar lift-off may contribute to eccentric polyethylene wear, particularly in designs of TKA which have flatter condyles. Coronal conformity is an important consideration in the design of a TKA.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Arthroplasty, Replacement, Knee*
  • Computer Simulation
  • Equipment Failure Analysis
  • Female
  • Femur / diagnostic imaging*
  • Humans
  • Imaging, Three-Dimensional
  • Knee Joint / diagnostic imaging*
  • Male
  • Middle Aged
  • Postoperative Complications / diagnostic imaging*
  • Prosthesis Design
  • Prosthesis Fitting
  • Radiography
  • Range of Motion, Articular / physiology
  • Tibia / diagnostic imaging
  • Video Recording
  • Weight-Bearing / physiology*