Biohazards assessment in large-scale zonal centrifugation

Appl Microbiol. 1975 Apr;29(4):484-90. doi: 10.1128/am.29.4.484-490.1975.

Abstract

A study was conducted to determine the biohazards associated with use of the large-scale zonal centrifuge for purification of moderate risk oncogenic viruses. To safely and conveniently assess the hazard, coliphage T3 was substituted for the virus in a typical processing procedure performed in a National Cancer Institute contract laboratory. Risk of personnel exposure was found to be minimal during optimal operation but definite potential for virus release from a number of centrifuge components during mechanical malfunction was shown by assay of surface, liquid, and air samples collected during the processing. High concentration of phage was detected in the turbine air exhaust and the seal coolant system when faulty seals were employed. The simulant virus was also found on both the centrifuge chamber interior and rotor surfaces.

MeSH terms

  • Air Microbiology
  • Centrifugation, Zonal* / instrumentation
  • Coliphages / isolation & purification
  • DNA Viruses
  • Environmental Exposure
  • Evaluation Studies as Topic
  • Humans
  • Laboratory Infection*
  • Models, Biological
  • Oncogenic Viruses / isolation & purification*
  • Safety