Assay of omeprazole and omeprazole sulfone by semi-microcolumn liquid chromatography with mixed-function precolumn

J Chromatogr B Biomed Sci Appl. 2001 Apr 25;754(2):487-93. doi: 10.1016/s0378-4347(01)00036-6.

Abstract

A column-switching system based on semi-microcolumns was used for direct analysis of omeprazole and omeprazole sulfone in human plasma samples. Plasma samples were injected into a mixed-function (MF Ph-1) column (35 mmx4.6 mm I.D.) to remove proteins and other non-specific peak producing substances from the analyte-containing time zone. The analyte-containing fraction was thereafter transferred to a C-18 semi-microcolumn (250 mmx1.5 mm I.D.) after concentration at the C-18 intermediate column. The absorbance at 302 nm in a ultraviolet (UV) detector was recorded to measure the concentration. The detection limit for omeprazole and omeprazole sulfone in the present method was 10 ng/ml. Interbatch variation (coefficient of variation) of the QC samples spanned less than 10% and intra-batch variation less than 2%. The recovery ratios of omeprazole and omeprazole sulfone were over 98%. The current method can be used as a simpler procedure with similar sensitivity and reproducibility as previously reported methods.

MeSH terms

  • Anti-Ulcer Agents / blood
  • Calibration
  • Chromatography, Liquid / methods*
  • Humans
  • Omeprazole / analogs & derivatives
  • Omeprazole / blood*
  • Omeprazole / metabolism
  • Quality Control
  • Reproducibility of Results