Comparison of durability of bioprostheses in tricuspid and mitral positions

Ann Thorac Surg. 2001 May;71(5 Suppl):S240-3. doi: 10.1016/s0003-4975(01)02557-7.

Abstract

Background: Few reports have assessed differences in the durability of mitral and tricuspid bioprostheses after simultaneous implantation of the same bioprosthesis in both positions. We investigated the long-term outcome after simultaneous mitral valve replacement (MVR) and tricuspid valve supraannular implantation (TVSI) with the Carpentier-Edwards bioprostheses in patients with severe tricuspid regurgitation and advanced mitral valve disease.

Methods: Between 1982 and 1998, 37 patients in our hospital underwent MVR and TVSI with Carpentier-Edwards bioprostheses. The mean age of the patients was 55+/-11 years. The average postoperative follow-up was 7.9+/-4.5 years after surgery (range 0 to 14.6 years, 315.1 patient-years). The follow-up rate was 100%. We evaluated the actuarial survival rate, the actuarial freedom from structural valve deterioration (SVD) and reoperation, and postoperative complications.

Results: The overall actuarial survival rate at 13 years after the operation was 69%+/-31%. The actuarial freedom from SVD and reoperation in the mitral and tricuspid positions were 78+/-22 and 100% and 70+/-30 and 90%+/-10% (p = 0.03), respectively. No patient had systemic or pulmonary thromboembolism, or complications associated with fatal arrhythmia.

Conclusions: These results suggest that the bioprostheses in the tricuspid position yield significantly better long-term results than those in the mitral position after simultaneous MVR and TVSI.

Publication types

  • Comparative Study

MeSH terms

  • Actuarial Analysis
  • Adult
  • Aged
  • Bioprosthesis*
  • Cause of Death
  • Female
  • Follow-Up Studies
  • Heart Valve Diseases / mortality
  • Heart Valve Diseases / surgery*
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Middle Aged
  • Mitral Valve / surgery*
  • Postoperative Complications / mortality
  • Prosthesis Design
  • Prosthesis Failure
  • Reoperation
  • Tricuspid Valve / surgery*