Simultaneous measurement of the trace elements Al, As, B, Be, Cd, Co, Cu, Fe, Li, Mn, Mo, Ni, Rb, Se, Sr, and Zn in human serum and their reference ranges by ICP-MS

Biol Trace Elem Res. 2001 Apr;80(1):77-93. doi: 10.1385/BTER:80:1:77.

Abstract

The goal of this article was to establish reference ranges of the concentration of trace elements in human serum and to compare these results with those reported by other authors. We describe the sample preparation and measurement conditions that allow the rapid, precise, and accurate determination of Al, As, B, Be, Cd, Co, Cu, Fe, Li, Mn, Mo, Ni, Rb, Se, Sr, and Zn in human serum samples (n = 110) by inductively coupled plasma-mass spectrometry (ICP-MS). Accuracy and precision were determined by analyzing three reconstituted reference serum samples by comparison with other methods and by the standard addition procedure. The advantages of the ICP-MS method include short time of analysis of the elements mentioned, low detection limit, high precision, and high accuracy. Disadventages include a high risk of contamination due to the presence of some of the elements of interest in the environment, the relatively delicate sample handling, and the high cost of the equipment.

MeSH terms

  • Adult
  • Calibration
  • Female
  • Humans
  • Indicators and Reagents
  • Male
  • Mass Spectrometry
  • Reference Values
  • Reproducibility of Results
  • Trace Elements / blood*

Substances

  • Indicators and Reagents
  • Trace Elements