Comparative Study
. 2001 Sep;18(5):340-2.
doi: 10.1136/emj.18.5.340.

Venous pH can safely replace arterial pH in the initial evaluation of patients in the emergency department

Affiliations
Comparative Study

Venous pH can safely replace arterial pH in the initial evaluation of patients in the emergency department

A M Kelly et al. Emerg Med J. 2001 Sep.

Abstract

Objective: This study aims to determine the extent of correlation of arterial and venous pH with a view to identifying whether venous samples can be used as an alternative to arterial values in the clinical management of selected patients in the emergency department.

Methods: This prospective study of patients who were deemed by their treating doctor to require an arterial blood gas analysis to determine their ventilatory or acid-base status, compared pH on an arterial and a venous sample taken as close to simultaneously as possible. Data were analysed using Pearson correlation and bias (Bland-Altman) methods.

Results: Two hundred and forty six patients were entered into the study; 196 with acute respiratory disease and 50 with suspected metabolic derangement. The values of pH on arterial and venous samples were highly correlated (r=0.92) with an average difference between the samples of -0.4 units. There was also a high level of agreement between the methods with the 95% limits of agreement being -0.11 to +0.04 units.

Conclusion: Venous pH estimation shows a high degree of correlation and agreement with the arterial value, with acceptably narrow 95% limits of agreement. Venous pH estimation is an acceptable substitute for arterial measurement and may reduce risks of complications both for patients and health care workers.

PubMed Disclaimer