Determination of trace amounts of total arsenic in environmental samples by hydride generation flow injection-AAS using a mixed acid as a pretreatment agent

Anal Sci. 2001 Sep;17(9):1067-71. doi: 10.2116/analsci.17.1067.

Abstract

A method is described for the determination of total arsenic by hydride generation-atomic absorption spectrophotometry using a mixed acid as a pretreatment. Hydride generation is done by the flow-injection method. The authors investigated in detail the temperature and time of decomposition using inorganic, organic arsenic and environmental standard samples, pretreated with nitric-perchloric-sulfuric mixed acid. By using a mixed acid as a pretreatment agent at 220 degrees C, the decomposition time could be shortened and the blank value of arsenic from the reagents used was reduced. The mixed acid of nitric-perchloric-sulfuric was also found to be effective as a pretreatment agent for organic arsenic compounds in which a dimethylated compound, sodium cacodylate or biological samples, is known to be one of the indecomposables. The present approach was proved to be satisfactory as a pretreatment for the quantitative analysis of trace amounts of total arsenic in liquid or solid environmental samples, such as geothermal water, sediments and biological samples.

MeSH terms

  • Acids
  • Arsenic / analysis*
  • Arsenicals / analysis*
  • Calibration
  • Environmental Pollutants / analysis*
  • Flow Injection Analysis
  • Indicators and Reagents
  • Reproducibility of Results
  • Solutions
  • Spectrophotometry, Atomic

Substances

  • Acids
  • Arsenicals
  • Environmental Pollutants
  • Indicators and Reagents
  • Solutions
  • Arsenic