A comparison of the lactate Pro, Accusport, Analox GM7 and Kodak Ektachem lactate analysers in normal, hot and humid conditions

Int J Sports Med. 2002 Feb;23(2):130-5. doi: 10.1055/s-2002-20133.

Abstract

This study aimed to compare the performance of a new portable lactate analyser against other standard laboratory methods in three conditions, normal (20 +/- 1.3 degrees C; 40 +/- 5 % RH), hot (40 +/- 2.5 degrees C; 40 +/- 5 % RH), and humid (20 +/- 1.1 degrees C; 82 +/- 6 % RH) conditions. Seven healthy males, ([Mean +/- SE]: age, 26.3 +/- 1.3 yr; height, 177.7 +/- 1.6 cm; weight, 77.4 +/- 0.9 kg, .VO(2)max, 56.1 +/- 1.9 ml x kg x min(-1)) undertook a maximal cycle ergometry test to exhaustion in the three conditions. Blood was taken every 3 min at the end of each stage and was analysed using the Lactate Pro LT-1710, the Accusport, the Analox GM7 and the Kodak Ektachem systems. The MANOVA (Analyser Type x Condition x Workload) indicated no interaction effect (F(42,660), = 0.45, p > 0.99, Power = 0.53). The data across all workloads indicated that the machines measured significantly differently to each other (F(4,743) = 14.652, p < 0.0001, Power = 1.00). The data were moderately to highly correlated. We conclude that the Lactate Pro is a simple and effective measurement device for taking blood lactate in a field or laboratory setting. However, we would caution against using this machine to compare data from other machines.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Analysis of Variance
  • Blood Chemical Analysis / instrumentation*
  • Hot Temperature
  • Humans
  • Humidity
  • Lactates / blood*
  • Male
  • Reference Values
  • Reproducibility of Results
  • Sports Medicine / instrumentation*

Substances

  • Lactates