Abstract
Plasma human immunodeficiency virus RNA in 491 clinical specimens was measured by LCx. There was a strong correlation with the results provided by other methods (r(2) values of 0.93 with Cobas-Monitor version 1.5 and of 0.95 with Quantiplex version 3.0). However, values were uniformly higher with LCx than with Quantiplex when non-B subtypes were tested.
Publication types
-
Comparative Study
-
Evaluation Study
-
Research Support, Non-U.S. Gov't
MeSH terms
-
HIV Infections / diagnosis*
-
HIV Infections / virology*
-
HIV-1 / genetics
-
HIV-1 / isolation & purification*
-
Humans
-
Ligase Chain Reaction / methods*
-
RNA, Viral / blood*
-
Reagent Kits, Diagnostic
-
Reproducibility of Results
-
Sensitivity and Specificity
-
Viral Load
Substances
-
RNA, Viral
-
Reagent Kits, Diagnostic