Accuracy of a narrow-diameter battery-powered endoscope in sedated and unsedated patients

Gastrointest Endosc. 2002 Apr;55(4):484-7. doi: 10.1067/mge.2002.122576.

Abstract

Background: Esophagoscopy with a portable battery-powered endoscope could provide a safe, inexpensive, and minimally invasive way to screen for Barrett's esophagus or esophageal varices. The use of such an instrument in an unsedated fashion has not been previously evaluated.

Methods: Patients referred for an EGD were recruited to undergo an additional examination with the battery-powered endoscope before EGD. In phase 1, (n = 42) patients received conscious sedation before the battery-powered endoscopic examination. In phase 2, (n = 56) patients were not sedated and were given the option of a peroral (n = 43) or transnasal (n = 13) endoscopy. Examiners were blinded to patient history and procedure indications. Esophageal findings, ease of intubation, optical quality, and patient comfort for the battery-powered endoscope and standard EGD were recorded by the endoscopist.

Results: Ninety-eight patients (60 men, 38 women, mean age 53 years) were recruited. The sensitivity for detecting Barrett's esophagus, esophageal tumors, and esophageal varices was 54.5%, 66.7%, and 80%, respectively. Ease of intubation and patient comfort as perceived by the endoscopist were not significantly different between the battery-powered endoscope and EGD. Optical quality was ranked as less than 4 (on a 5-point scale with 5 = standard EGD and 1 = poor) in 42% of battery-powered endoscopic examinations. There were no complications.

Conclusion: The accuracy of esophageal examination with a 3.1-mm endoscope is substantially inferior to standard EGD. Thus, the battery-powered endoscope would not be useful for screening patients for Barrett's esophagus or varices unless improvements in optical quality and visualization are made.

Publication types

  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Conscious Sedation
  • Esophageal Diseases / diagnosis
  • Esophagoscopes*
  • Female
  • Fiber Optic Technology
  • Humans
  • Male
  • Middle Aged
  • Sensitivity and Specificity