Correction of slab boundary artifact using histogram matching

J Magn Reson Imaging. 2002 May;15(5):610-7. doi: 10.1002/jmri.10094.

Abstract

Purpose: To improve the imaging efficiency of multi-slab three-dimensional acquisitions by decreasing slab overlap and the number of blanked slices.

Materials and methods: Application of the histogram matching technique for correction of the slab boundary artifact (SBA) due to non-ideal radiofrequency pulse profiles has been proposed. The technique was validated by studying its effectiveness in reducing the SBA in phantom images, as well as in proton density- and T2-weighted brain images.

Results: The correction algorithm considerably decreases the disruption of image volume continuity due to the SBA, and improves slice-to-slice intensity continuity.

Conclusion: The proposed technique can greatly reduce the SBA when the SBA can be described as a global intensity variation along the slab. The technique may be used to improve imaging efficiency of the multi-slab three-dimensional fast spin-echo sequences by decreasing the amount of required slab overlap.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Algorithms*
  • Artifacts
  • Brain / anatomy & histology*
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging*
  • Phantoms, Imaging