Validation of diffusion tensor MRI-based muscle fiber tracking

Magn Reson Med. 2002 Jul;48(1):97-104. doi: 10.1002/mrm.10198.

Abstract

Diffusion-tensor (DT) MRI fiber tracking may potentially be used for in vivo structural analysis. The purpose of this study was to assess quantitatively the ability of a DT-MRI fiber-tracking algorithm to measure the fiber orientation (pennation) in skeletal muscle in vivo. In five adult Sprague-Dawley rats, the pennation angle (theta) was measured in the rat lateral gastrocnemius with DT-MRI (theta(DT-MRI)) and by direct anatomical inspection (DAI) (theta(DAI)). The mean theta(DT-MRI) was not significantly different from the mean theta(DAI). In addition, the two methods were highly correlated (r = 0.89) and the regression of theta(DT-MRI) on theta(DAI) resulted in a slope not significantly different from 1 and an intercept not significantly different from zero. These data indicate that DT-MRI-based fiber tracking as implemented here is a valid tool for in vivo structural analysis of small-animal skeletal muscle.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Magnetic Resonance Imaging / methods*
  • Muscle Fibers, Skeletal / cytology*
  • Muscle, Skeletal / cytology*
  • Rats