Are TrueFISP images T2/T1-weighted?

Magn Reson Med. 2002 Oct;48(4):684-8. doi: 10.1002/mrm.10260.

Abstract

Images acquired using the TrueFISP technique (true fast imaging with steady-state precession) are generally believed to exhibit T(2)/T(1)-weighting. In this study, it is demonstrated that with the widely used half-flip-angle preparation scheme, approaching the steady state requires a time length comparable to the scan time such that the transient-state response may dominate the TrueFISP image contrast. Two-dimensional images of the human brain were obtained using various phase-encoding matrices to investigate the transient-state signal behavior. Contrast between gray and white matter was found to change significantly from proton-density- to T(2)/T(1)-weighted as the phase-encoding matrix size increased, which was in good agreement with theoretical predictions. It is concluded that TrueFISP images in general exhibit T(2)/T(1)-contrast, but should be more appropriately regarded as exhibiting a transient-state combination of proton-density and T(2)/T(1) contrast under particular imaging conditions. Interpretation of tissue characteristics from TrueFISP images in clinical practice thus needs to be exercised with caution.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Brain / anatomy & histology*
  • Humans
  • Magnetic Resonance Imaging / methods*