Effective radiation dose from semicoronal CT of the sacroiliac joints in comparison with axial CT and conventional radiography

Eur Radiol. 2002 Nov;12(11):2820-5. doi: 10.1007/s00330-001-1297-7. Epub 2002 Feb 16.

Abstract

The aim of this study was to evaluate the radiation dose given by semicoronal CT of the sacroiliac joints (SIJs) in comparison with axial CT and conventional radiography. The total effective radiation doses given by serial contiguous semicoronal and axial CT, using 5-mm slices, 120 kV and 330 mAs, were determined by measurement of organ doses using an anthropomorphic Rando Alderson phantom paced with thermoluminescence dosimeters. The doses given by conventional antero-posterior (AP) and oblique projections of the SIJs were determined similarly. In a female the total effective dose by semicoronal CT was found to be more than six times lower than by axial CT and 2.5 times lower than the dose use to obtain a conventional AP radiograph, the values being 102, 678, and 255 micro Sv, respectively. The effective dose by semicoronal CT was only a little higher than the dose given to obtain two oblique radiographs. In a male with lead protection of the gonads the dose by semicoronal CT was four times lower than by axial CT, but higher than by conventional radiography. In conclusion, the effective dose by semicoronal CT of the SIJs is lower than by axial CT, and in females a semicoronal CT implies a lower effective radiation dose that used to obtain an AP radiograph.

MeSH terms

  • Female
  • Humans
  • Male
  • Phantoms, Imaging
  • Radiation Dosage*
  • Sacroiliac Joint / diagnostic imaging*
  • Sacroiliac Joint / radiation effects
  • Tomography, X-Ray Computed*