Miniature cylindrical ion trap mass spectrometer

Anal Chem. 2002 Dec 15;74(24):6145-53. doi: 10.1021/ac020494d.

Abstract

A miniature cylindrical ion trap mass spectrometer is described, and preliminary data are presented. Functionality and performance of laboratory-scale instruments have been maintained to the extent possible in this battery-operated mass spectrometer. Capabilities include tandem mass spectrometry experiments. Custom-designed electronic components include the RF scanning and amplification system, data acquisition components, and lens power supplies, as well as a custom-software application. Direct leak and membrane introduction inlet systems are used for sample introduction. A mass/charge range of approximately 250 Th with unit mass resolution has been demonstrated.