Effect of limb cooling on peripheral and global oxygen consumption in neonates
- PMID: 12598504
- PMCID: PMC1721525
- DOI: 10.1136/fn.88.2.f139
Effect of limb cooling on peripheral and global oxygen consumption in neonates
Abstract
Aim: To evaluate peripheral oxygen consumption (VO(2)) measurements using near infrared spectroscopy (NIRS) with arterial occlusion in healthy term neonates by studying the effect of limb cooling on peripheral and global VO(2).
Subjects and methods: Twenty two healthy term neonates were studied. Peripheral VO(2) was measured by NIRS using arterial occlusion and measurement of the oxyhaemoglobin (HbO(2)) decrement slope. Global VO(2) was measured by open circuit calorimetry. Global and peripheral VO(2) was measured in each neonate before and after limb cooling.
Results: In 10 neonates, a fall in forearm temperature of 2.2 degrees C (mild cooling) decreased forearm VO(2) by 19.6% (p < 0.01). Global VO(2) did not change. In 12 neonates, a fall in forearm temperature of 4 degrees C (moderate cooling) decreased forearm VO(2) by 34.7% (p < 0.01). Global VO(2) increased by 17.6% (p < 0.05).
Conclusions: The NIRS arterial occlusion method is able to measure changes in peripheral VO(2) induced by limb cooling. The changes are more pronounced with moderate limb cooling when a concomitant rise in global VO(2) is observed. Change in peripheral temperature must be taken into consideration in the interpretation of peripheral VO(2) measurements in neonates.
Similar articles
-
Effect of a change in global metabolic rate on peripheral oxygen consumption in neonates.Arch Dis Child Fetal Neonatal Ed. 2003 Mar;88(2):F143-6. doi: 10.1136/fn.88.2.f143. Arch Dis Child Fetal Neonatal Ed. 2003. PMID: 12598505 Free PMC article.
-
Forearm and calf tissue oxygenation in term neonates measured with near-infrared spectroscopy.J Physiol Sci. 2007 Oct;57(5):317-9. doi: 10.2170/physiolsci.SC004407. Epub 2007 Oct 6. J Physiol Sci. 2007. PMID: 17916280
-
Measurement of peripheral oxygen utilisation in neonates using near infrared spectroscopy: comparison between arterial and venous occlusion methods.Early Hum Dev. 2000 Mar;57(3):211-24. doi: 10.1016/s0378-3782(99)00076-6. Early Hum Dev. 2000. PMID: 10742611
-
Peripheral Muscle Near-Infrared Spectroscopy in Neonates: Ready for Clinical Use? A Systematic Qualitative Review of the Literature.Neonatology. 2015;108(4):233-45. doi: 10.1159/000433515. Epub 2015 Aug 25. Neonatology. 2015. PMID: 26338668 Review.
-
Peripheral oxygenation and management in the perinatal period.Semin Fetal Neonatal Med. 2010 Aug;15(4):208-15. doi: 10.1016/j.siny.2010.03.005. Epub 2010 May 20. Semin Fetal Neonatal Med. 2010. PMID: 20493790 Review.
Cited by
-
Peripheral fractional oxygen extraction measured with near-infrared spectroscopy in neonates-A systematic qualitative review.Front Pediatr. 2022 Aug 23;10:940915. doi: 10.3389/fped.2022.940915. eCollection 2022. Front Pediatr. 2022. PMID: 36081622 Free PMC article.
-
The Predictive Value of Pulse Oximeters for Pulse Improvement after Angiography in Infants and Children.Iran J Pediatr. 2016 Jul 10;26(5):e5833. doi: 10.5812/ijp.5833. eCollection 2016 Oct. Iran J Pediatr. 2016. PMID: 28203338 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources