New transcutaneous jaundice device with two optical paths

J Perinat Med. 2003;31(1):81-8. doi: 10.1515/JPM.2003.012.

Abstract

Objectives: To confirm the accuracy and precision of transcutaneous bilirubin (TcB) values measured by a new device with two optical paths (JM-103) and the value of total serum bilirubin (TSB) level in clinical units of measurement.

Methods: For comparison of the levels of accuracy and precision of JM-103 and the old device (JM-102), serum samples were collected from 77 Japanese infants in three different hospitals including 24 preterm infants and 53 term infants. Measurement of TcB by JM-103 and JM-102 were performed on the forehead of each infant within 30 min before or after blood sampling.

Results: The range of TSB was limited to 19.6 mg/dL and to 17 mg/dL for preterm infants. The correlation coefficients for all subjects (r = 0.94) and for term and preterm subjects between TcB measured by JM-103 and TSB was higher than that between TcB measured by JM-102 and TSB. The regression line in term infants between TcB measured by JM-103 and TcB was similar to that in preterm infants. The error distribution of TcB measured by JM-102 and TSB for all subjects (0.00 +/- 2.21) and for term and preterm subjects was larger than that of TcB measured by JM-103 and TSB (all subjects, 0.30 +/- 1.55).

Publication types

  • Clinical Trial
  • Comparative Study

MeSH terms

  • Bilirubin / analysis*
  • Chromatography, High Pressure Liquid
  • Color
  • Equipment Design
  • Gestational Age
  • Humans
  • Infant, Newborn
  • Infant, Premature / blood
  • Jaundice, Neonatal / diagnosis*
  • Jaundice, Neonatal / metabolism
  • Neonatal Screening / instrumentation*
  • Predictive Value of Tests
  • Reference Values
  • Regression Analysis
  • Subcutaneous Tissue / metabolism*

Substances

  • Bilirubin