Performance evaluation of a hand-held, semiconductor (CdZnTe)-based gamma camera
- PMID: 12677308
- DOI: 10.1007/s00259-002-1067-z
Performance evaluation of a hand-held, semiconductor (CdZnTe)-based gamma camera
Abstract
We have designed and developed a small field of view gamma camera, the eZ SCOPE, based on use of a CdZnTe semiconductor. This device utilises proprietary signal processing technology and an interface to a computer-based imaging system. The purpose of this study was to evaluate the performance of the eZ scope in comparison with currently employed gamma camera technology. The detector is a single wafer of 5-mm-thick CdZnTe that is divided into a 16x16 array (256 pixels). The sensitive area of the detector is a square of dimension 3.2 cm. Two parallel-hole collimators are provided with the system and have a matching (256 hole) pattern to the CdZnTe detector array: a low-energy, high-resolution parallel-hole (LEHR) collimator fabricated of lead and a low-energy, high-sensitivity parallel-hole (LEHS) collimator fabricated of tungsten. Performance measurements and the data analysis were done according to the procedures of the NEMA standard. We also studied the long-term stability of the system with continuous use and variations in ambient temperature. Results were as follows. INTRINSIC ENERGY RESOLUTION: 8.6% FWHM at 141 keV.LINEARITY: There was excellent linearity between the observed photopeaks and the known gamma ray energies for the given isotopes. INTRINSIC SYSTEM UNIFORMITY: For the central field of view, the integral uniformity and the differential uniformity were, respectively, 1.6% and 1.3% with the LEHR collimator and 1.9% and 1.2% with the LEHS collimator. SYSTEM SPATIAL RESOLUTION: The FWHM measurements made at the surface of the collimator were 2.2 mm (LEHR) and 2.9 mm (LEHS).CONTRAST TEST: The average S/N ratios (i.e. counts in the irradiated pixel divided by counts in the surrounding pixels) for the inner ring pixels (8)/outer ring pixels (16) using the LEHS collimator and LEHR collimator were 3.2%/0.2% and 3.7%/0.3%, respectively. COUNT RATE CHARACTERISTICS: We could not determine the maximum count rate and the 20% loss count rate from these data because the plateau was not reached while using the solutions measured. SYSTEM SENSITIVITY: The average acquisitions were 11,052 cpm/MBq (LEHR) and 28,590 cpm/MBq (LEHS). TEMPERATURE DEPENDENCE: The system displayed minimum corresponding shift in cps with temperature changes in the measured temperature range. We designed and developed a semiconductor-based gamma camera using CdZnTe. The basic performance of this camera compares favourably with the existing gamma camera technology that is deployed in the medical field today. The most significant differences include the spatial resolution, sensitivity, high count rate characteristics and energy resolution. We believe that this device will be of value for a number of clinical applications including sentinel node detection and radiopharmaceutical-guided surgery.
Similar articles
-
[Evaluation of efficiency of a semiconductor gamma camera].Kaku Igaku. 2002 Nov;39(4):549-53. Kaku Igaku. 2002. PMID: 12607244 Japanese.
-
Performance characteristics of a new pixelated portable gamma camera.Med Phys. 2012 Jun;39(6):3435-44. doi: 10.1118/1.4718874. Med Phys. 2012. PMID: 22755723 Free PMC article.
-
Evaluation of a CdTe semiconductor based compact γ camera for sentinel lymph node imaging.Med Phys. 2011 Mar;38(3):1547-60. doi: 10.1118/1.3555034. Med Phys. 2011. PMID: 21520866
-
Nuclear probes and intraoperative gamma cameras.Semin Nucl Med. 2011 May;41(3):166-81. doi: 10.1053/j.semnuclmed.2010.12.004. Semin Nucl Med. 2011. PMID: 21440694 Review.
-
Improvements in SPECT technology for cerebral imaging.Semin Nucl Med. 1985 Oct;15(4):335-46. doi: 10.1016/s0001-2998(85)80012-x. Semin Nucl Med. 1985. PMID: 3904004 Review.
Cited by
-
Evaluation of sentinel lymph node localization in malignant melanoma by preoperative semiconductor gamma camera and planar lymphoscintigraphy.J Appl Clin Med Phys. 2023 Aug;24(8):e14077. doi: 10.1002/acm2.14077. Epub 2023 Jun 26. J Appl Clin Med Phys. 2023. PMID: 37357570 Free PMC article.
-
Nuclear-medicine probes: Where we are and where we are going.Med Phys. 2022 Jul;49(7):4372-4390. doi: 10.1002/mp.15690. Epub 2022 May 20. Med Phys. 2022. PMID: 35526220 Free PMC article. Review.
-
Above and Beyond Robotic Surgery and 3D Modelling in Paediatric Cancer Surgery.Front Pediatr. 2021 Dec 20;9:777840. doi: 10.3389/fped.2021.777840. eCollection 2021. Front Pediatr. 2021. PMID: 34988038 Free PMC article. Review.
-
Characterisation of a hand-held CZT-based gamma camera for 177Lu imaging.EJNMMI Phys. 2020 Jul 8;7(1):46. doi: 10.1186/s40658-020-00313-y. EJNMMI Phys. 2020. PMID: 32642884 Free PMC article.
-
Intraoperative biophotonic imaging systems for image-guided interventions.Nanophotonics. 2019 Jan;8(1):99-116. doi: 10.1515/nanoph-2018-0134. Epub 2018 Dec 14. Nanophotonics. 2019. PMID: 31187017 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources