Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?
- PMID: 12798581
- DOI: 10.1016/s0735-1097(03)00424-8
Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?
Abstract
Objectives: The aim of this study was to determine whether airport metal detector gates (AMDGs) interfere with pacemakers (PMs) or implantable cardioverter-defibrillators (ICDs).
Background: It is currently unknown whether AMDGs interfere with implanted PMs or ICDs.
Methods: A total of 348 consecutive patients (200 PM and 148 ICD recipients) have been tested for the occurrence of electromagnetic interference (EMI) within the electromagnetic field of a worldwide-used airport metal detector.
Results: No interference, such as pacing or sensing abnormalities, was observed in any of the 200 PM and 148 ICD patients; also no reprogramming occurred.
Conclusions: In vivo testing of PM and ICD systems showed no EMI with a standard AMDG. Clinically relevant interactions with implanted PMs or ICDs seem unlikely.
Similar articles
-
Electromagnetic interference with cardiac pacemakers and implantable cardioverter-defibrillators from low-frequency electromagnetic fields in vivo.Europace. 2013 Mar;15(3):388-94. doi: 10.1093/europace/eus345. Epub 2012 Nov 1. Europace. 2013. PMID: 23125355
-
In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillators.Heart Rhythm. 2010 Jan;7(1):99-107. doi: 10.1016/j.hrthm.2009.09.071. Epub 2009 Oct 12. Heart Rhythm. 2010. PMID: 20129290
-
Safety of screening procedures with hand-held metal detectors among patients with implanted cardiac rhythm devices: a cross-sectional analysis.Ann Intern Med. 2011 Nov 1;155(9):587-92. doi: 10.7326/0003-4819-155-9-201111010-00005. Ann Intern Med. 2011. PMID: 22041947
-
Implantable rhythm devices and electromagnetic interference: myth or reality?Expert Rev Cardiovasc Ther. 2008 Jul;6(6):823-32. doi: 10.1586/14779072.6.6.823. Expert Rev Cardiovasc Ther. 2008. PMID: 18570620 Review.
-
Electromagnetic interference in cardiac rhythm management devices.AACN Clin Issues. 2004 Jul-Sep;15(3):391-403. doi: 10.1097/00044067-200407000-00007. AACN Clin Issues. 2004. PMID: 15475813 Review.
Cited by
-
Hearts in the sky: understanding the cardiovascular implications of air travel.Future Cardiol. 2024;20(11-12):651-660. doi: 10.1080/14796678.2024.2396257. Epub 2024 Sep 20. Future Cardiol. 2024. PMID: 39301896 Review.
-
[Electromagnetic interference : Pacemakers, cardiac resynchronization therapy devices, implantable cardioverter-defibrillator].Herzschrittmacherther Elektrophysiol. 2022 Sep;33(3):297-304. doi: 10.1007/s00399-022-00875-7. Epub 2022 Jul 4. Herzschrittmacherther Elektrophysiol. 2022. PMID: 35781834 Review. German.
-
Travelling with heart failure: risk assessment and practical recommendations.Nat Rev Cardiol. 2022 May;19(5):302-313. doi: 10.1038/s41569-021-00643-z. Epub 2022 Jan 6. Nat Rev Cardiol. 2022. PMID: 34992256 Review.
-
Is air travel safe for patients with cardiac implantable electronic devices?Anatol J Cardiol. 2021 Aug;25(Suppl 1):26-28. doi: 10.5152/AnatolJCardiol.2021.S110. Anatol J Cardiol. 2021. PMID: 34464297 Free PMC article.
-
Commercial Air Travel for Passengers With Cardiovascular Disease: Stressors of Flight and Aeromedical Impact.Curr Probl Cardiol. 2021 Mar;46(3):100746. doi: 10.1016/j.cpcardiol.2020.100746. Epub 2020 Nov 2. Curr Probl Cardiol. 2021. PMID: 33213942 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical