Evaluation of the transmitted exposure through lead equivalent aprons used in a radiology department, including the contribution from backscatter

Med Phys. 2003 Jun;30(6):1033-8. doi: 10.1118/1.1573207.

Abstract

A study was conducted to evaluate the radiation transmission through lead equivalent aprons that are used in a radiology department. A large area beam (poor geometry) was employed for the transmission measurements, and backscatter was simulated by placing 7" of Lucite behind each apron. Separate ionization chambers were used to measure the incident and transmitted x-ray beams. Transmission measurements were made at 70 kVp and 100 kVp through aprons and protective shields from eight different vendors that were marked 0.25 mm and 0.5 mm lead equivalent. Transmissions through 0.254 mm and 0.508 mm of pure lead were also measured and were compared with the transmissions through the lead equivalent materials. In addition, the area densities of the aprons were measured to compare radiation transmission with respect to the weights of the aprons. At 70 kVp, the transmission through 0.254 mm of pure lead was 5.4% and the transmissions through the 0.25 mm lead equivalent materials were 4.3% to 10.2% with a mean value of 7.1% and a standard deviation (s.d.) of 1.4%. At 100 kVp, the values were 15% for 0.254 mm pure lead and 12.3% to 20.7% (mean 16.8%, s.d. 2.1%) for the 0.25 mm lead equivalent materials. The transmission through the 0.508 mm pure lead sample was 0.9% at 70 kVp, and the corresponding transmissions through the 0.5 mm lead equivalent materials were 0.6% to 1.6% (mean 1.0%, s.d. 0.2%). At 100 kVp, the transmission through the 0.508 mm lead sample was 5% and those through the 0.5 mm lead equivalent materials were 3.5% to 6.7% (mean 4.9%, s.d. 0.7%). The radiation transmissions at 70 kVp, through two "lead-free" 0.5 mm lead equivalent aprons, were 1.7% and 1.9% and at 100 kVp the transmissions were 6.1% and 6.8%, respectively. This study indicates that there is a need to establish methods for acceptance testing of aprons and a need to establish acceptance limits for the x-ray transmission of aprons at specific kVp values. There is also a need for the establishment of appropriate methods and frequencies of routine quality assurance testing of radiation protection aprons.

Publication types

  • Comparative Study
  • Evaluation Study
  • Validation Study

MeSH terms

  • Equipment Failure Analysis / methods*
  • Lead / radiation effects*
  • Occupational Exposure / analysis*
  • Protective Clothing*
  • Radiation Dosage*
  • Radiation Protection / instrumentation*
  • Radiation Protection / methods*
  • Radiology Department, Hospital
  • Radiometry / methods
  • Reproducibility of Results
  • Scattering, Radiation
  • Sensitivity and Specificity
  • X-Rays*

Substances

  • Lead