A study on radiation doses and irradiated areas in cerebral embolisation

Br J Radiol. 2003 Aug;76(908):546-52. doi: 10.1259/bjr/26353198.

Abstract

Patient radiation doses during interventional radiology procedures may reach the thresholds for radiation-induced skin and eye lens injuries. This study investigates the irradiated areas and doses received by patients undergoing cerebral embolisation, which is regarded as a high dose interventional radiology procedure. For each procedure the fluoroscopic and digital dose-area product (DAP), the fluoroscopic time, the total number of acquired images and entrance-skin dose (ESD) calculated by the angiographic unit were recorded. The ESD was measured by means of thermoluminescent dosimeters. In this study, the skin, eye and thyroid gland doses and the irradiated area for 30 patients were recorded. The average ESD was found to be 0.77 Gy for the posteroanterior plane and 0.78 Gy for the lateral plane. The average DAP was 48 Gy cm(2) for the posteroanterior plane and 58 Gy cm(2) for the lateral plane. The patient's average right eye dose was 60 mGy and the dose to the thyroid gland was 24 mGy. Seven patients received a dose above 1 Gy, one patient exceeded the threshold for transient erythema and one exceeded the threshold for temporary epilation. A good correlation between the DAP and the ESD for both planes has been found. The doctor's eye dose has also been measured for 17 procedures and the average dose per procedure was 0.13 mGy.

MeSH terms

  • Embolization, Therapeutic / methods*
  • Female
  • Fluoroscopy
  • Humans
  • Intracranial Aneurysm / therapy*
  • Male
  • Radiation Dosage
  • Radiography, Interventional*