Assessment of a wrist-worn device in the detection of obstructive sleep apnea

Sleep Med. 2003 Sep;4(5):435-42. doi: 10.1016/s1389-9457(03)00111-4.

Abstract

Objective: To assess the accuracy of a wrist-worn device (Watch_PAT100) to diagnose obstructive sleep apnea (OSA).

Methods: Thirty adult subjects with and without suspected OSA simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the Watch_PAT100 during a full-night recording. PSG sleep and respiratory events were scored according to standard criteria. Watch_PAT data were analyzed with an automated computerized algorithm which calculated the frequency of respiratory events per hour of actigraphy measured sleep using a combination of peripheral arterial tonometry (PAT) signal attenuation, desaturation on pulse oximetry, and changes in heart rate. This yielded a PAT apnea hypopnea index (AHI).

Results: Mean age was 47.0+/-14.8 years, mean body mass index 31.0+/-7.6 kg/m(2), mean PSG AHI 23+/-23.9 events per hour, and mean PAT AHI 23+/-15.9 events per hour. There was a significant correlation between PAT AHI and AHI by PSG (r=0.87, P<0.001). To assess sensitivity and specificity of Watch_PAT, we constructed receiver operator characteristic curves using a variety of AHI threshold values (10, 15, 20, and 30 events per hour). Optimal combinations of sensitivity and specificity for the various thresholds were 82.6/71.4, 93.3/73.3, 90.9/84.2, and 83.3/91.7, respectively.

Conclusions: The Watch_PAT is a device that can detect OSA with reasonable accuracy. Thus, the Watch_PAT may be a useful method to diagnose OSA.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • Case-Control Studies
  • Equipment Design
  • Equipment and Supplies*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Polysomnography
  • ROC Curve
  • Sensitivity and Specificity
  • Sleep Apnea, Obstructive / diagnosis*
  • Wrist*