A prospective evaluation of interrupted nitinol surgical clips in arteriovenous fistula for hemodialysis

Am J Surg. 2003 Dec;186(6):625-30. doi: 10.1016/j.amjsurg.2003.08.007.

Abstract

Background: The use of a self-closing nitinol surgical clip (Coalescent Surgical U-Clip) is a novel technique of creating an interrupted vascular anastomosis, which also eliminates the knot tying associated with a conventional vascular anastomosis. The purpose of this study was to compare the clinical outcome of arteriovenous fistulae (AVF) for hemodialysis as constructed using interrupted U-Clip devices and conventional continuous polypropylene sutures.

Methods: We prospectively studied the clinical course of 132 patients (149 access) undergoing either forearm autologous arteriovenous fistula (FAVF, n = 69) or upper arm autologous arteriovenous fistula (UAVF, n = 80) or creations during a 39-month period. Among the FAVF, U-Clips and polypropylene sutures were used in 37 and 32 anastomoses, respectively. Among the UAVF, U-Clips and polypropylene sutures were used in 41 and 30 anastomoses, respectively. Primary patency, complication, and maturation rates were analyzed.

Results: Clipped FAVF had an improved maturation rate at 6 weeks (32 of 37, 86%) when compared with sutured FAVF (22 of 32, 69%, P <0.05). There was no difference in the maturation rate among UAVF constructed with either U-Clips or sutures (88% versus 87%, not significant). The primary patency rates at 12, 24, and 36 months were 91%, 84%, and 75% for the clipped FAVF; and 83%, 74%, and 61% for the sutured FAVF (P <0.05) There was no difference in the patency rate of UAVF constructed with either U-Clips or polypropylene sutures at either 12, 24 or 36 moths (91%, 80%, and 75% versus 83%, 77%, and 69%, respectively).

Conclusions: The U-Clips are a viable alternative to sutures for creating vascular anastomosis. It provides a improved maturation and patency rates when compared with the conventional sutured AVF in the forearm. The clinical benefit of the U-Clips may be due in part to the improved compliance created by the interrupted anastomotic technique.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adult
  • Alloys*
  • Anastomosis, Surgical / instrumentation*
  • Anastomosis, Surgical / methods
  • Arm / blood supply
  • Arteriovenous Shunt, Surgical*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Polypropylenes
  • Prospective Studies
  • Renal Dialysis*
  • Surgical Instruments*
  • Survival Analysis
  • Sutures
  • Vascular Patency

Substances

  • Alloys
  • Polypropylenes
  • nitinol