It has been shown that preheating the femoral stem prior to insertion minimizes interfacial porosity at the stem-cement interface. In this study, the effects of methylmethacrylate monomer temperature prior to mixing on the properties of stem-cement interface and cement polymerization were evaluated for 4 degrees C, room temperature, and 37 degrees C using a test model and cementing techniques that simulated a clinical situation. The nature and extent of interfacial porosity of stem-cement interface was quantified, the static shear strength of the stem-cement interface determined, and the time and temperature of polymerization at the cement-bone interface were measured. Compared to RT monomer, preheating monomer to 37 degrees C produced higher polymerization temperatures and greater initial interfacial shear strength with an unchanged amount of interfacial porosity. Precooling monomer to 4 degrees C produced lower polymerization temperatures and decreased initial interfacial shear strength, with the amount of interfacial porosity unchanged compared to the RT group. Although clinical techniques of preheating or precooling bone cement have some effects on the properties of the stem-cement interface and cement polymerization, they do not appear to enhance implant fixation.
Copyright 2003 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 68B: 186-190, 2004