Measurement of a 103Pd solution using the TDCR method by LSC

Appl Radiat Isot. 2004 Feb-Apr;60(2-4):439-45. doi: 10.1016/j.apradiso.2003.11.056.

Abstract

Palladium 103 is a radionuclide used in brachytherapy sources for the treatment of prostate cancers and also for other medical applications. It decays through electron capture to excited levels of 103Rh. This paper describes the calculation method used to compute the detection efficiency in the framework of the triple to double coincidence ratio model. The calculation of the energy transferred to the scintillator is made by considering the various atomic events following the electron capture and the electron conversion. The energy deposited in the scintillator after the absorption of X-rays is calculated using the PENELOPE stochastic calculation code. The main contributors to the final uncertainty and their covariance matrix are discussed. As the calculation method cannot be reduced to an explicit function, this paper describes the Monte Carlo method used for the evaluation of uncertainties.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Algorithms
  • Beta Particles
  • Brachytherapy / standards*
  • Calibration / standards*
  • Equipment Design
  • Equipment Failure Analysis
  • France
  • Palladium / analysis*
  • Palladium / standards*
  • Radioisotopes / analysis*
  • Radioisotopes / standards*
  • Radiopharmaceuticals / analysis
  • Radiopharmaceuticals / standards
  • Reference Standards
  • Reproducibility of Results
  • Scintillation Counting / instrumentation
  • Scintillation Counting / methods*
  • Scintillation Counting / standards*
  • Sensitivity and Specificity
  • Solutions
  • Spectrometry, Gamma / instrumentation
  • Spectrometry, Gamma / methods
  • Spectrometry, Gamma / standards

Substances

  • Radioisotopes
  • Radiopharmaceuticals
  • Solutions
  • Palladium